A retrospective study evaluated whether reducing scan time at specific imaging time points can preserve absorbed dose accuracy during lutetium-177 dotatate therapy. Using a ring-shaped CZT SPECT/CT system, investigators compared shortened protocols against full 28-minute reference acquisitions at 4 hours, 24 hours, and 48–72 hours post-injection. The results suggest that timing matters: the best kidney dosimetry accuracy came from reducing scan time only at the earliest 4-hour time point (protocol A4), while tumor dosimetry performed best when shortening was applied at the 24-hour stage (protocol A2). The paper reports root mean square and mean absolute percentage errors within defined ranges for both organs. If validated prospectively, the approach could reduce patient burden and throughput constraints in radiopharmaceutical imaging without compromising dosimetry for treatment planning.